Accu 您所在的位置:网站首页 accu chek test solution Accu

Accu

#Accu| 来源: 网络整理| 查看: 265

A meter or test strip error has occurred. Repeat the blood glucose test. If a second E-3 code appears, perform a control test with the control solution and a new test strip.

If the control result is within the acceptable range, review the proper testing procedure and repeat the blood glucose test with a new test strip. If the control result is not within the acceptable range, see the Understanding Out-of-Range Control Results section in the chapter Control Tests.

In rare cases, an E-3 error code may indicate that your blood glucose is extremely high and above the system’s reading range. If you do not feel like your blood glucose is extremely high, repeat the blood glucose test. See the Unusual Blood Glucose Results section in the chapter Blood Glucose Tests. If the E-3 code still appears for your blood glucose test, contact your healthcare professional immediately.